3. Initial and routine in process control of the BioBLU® 0.3c
during cultivation.
(a)
Following the attachment phase, transfer the bioreactor
to the biosafety cabinet, remove the remaining port lid as
previously described, resuspend the MCs using a 10 mL
sterile pipette, and take a homogenous 7 mL sample (see
Note 18). Subsequent sample treatment prior to the
analyses is more closely described in Subheading 3.4.
(b)
Transfer the bioreactor back to the control unit and reac-
tivate all the control loops to maintain the predefined
process conditions, such as the Ns1u criteria for the
remainder of the cultivation.
(c)
Subsequent routine sampling (5 mL) should be per-
formed every 24 h post-inoculation, especially prior to
and following the media exchange (see Fig. 3).
4. Perform a 50% media exchange 4 d post-inoculation to prevent
the accumulation of inhibitory cell metabolites and the glucose
concentration
from
dropping
below
5.55
mmol
L1
(¼ 1 g L1).
(a)
Transfer the bioreactor to the biosafety cabinet and allow
the MCs to sediment.
(b)
Open the remaining port lid as previously described and
remove 50% of the supernatant (approximately 61.5 mL)
while ensuring that no MCs are aspirated, and replace with
fresh pre-warmed culture media (see Note 6 and 13).
(c)
Following the media exchange, refasten the sterile port lid
and return the bioreactor to the control unit and reacti-
vate all the control loops.
5. The harvest criterion is met at least 24 h prior to reaching the
stationary phase of cell growth (within 5–6 days post-
inoculation), at which point a final harvest may be performed.
(a)
Deactivate all the control loops and allow the MCs to
sediment, then transfer the bioreactor to the biosafety
cabinet and remove the remaining port lid of the bio-
reactor as previously described.
(b)
Proceed to remove the supernatant without aspirating
MCs and replace with pre-warmed DPBS. Allow the
MCs
to
sediment,
then
replace
the
DPBS
with
pre-warmed TrypLE Select 1 (see Note 13).
(c)
Refasten the sterile port lid and return the bioreactor to
the control unit. Reattach the necessary instruments and
periphery, reactivate all the control loops and set the
agitation rate to Ns1u for 15 min (see Note 14).
Mesenchymal Stem Cell Expansion at Benchtop-Scale
99